Escort 3016B User Manual
Page 116
15
Ethernet interface
Page 18
14.5.11.
Launching plotting and recording
14.5.12.
Diagrams
HEADER
PARAMETERS
EXAMPLES
RECord
P1
RECORD ON
Start or Stop
of plotting in DIRECT mode,
or recording in other modes.
P1= ON : start
OFF : stop
TRIG: forcing of trigger
TRIGREC : forcing of memory trigger in DIRECT mode
RECord ?
Returns the command and the ratio of the MEMORY recording
In DIRECT mode, the plotting will began
after start trigger is true
The triggering can be forced with
RECORD TRIG for start and RECORD
OFF for stop
HEADER
PARAMETERS
EXAMPLES
GRID
P1,P2
Defines the diagrams
P1 = Number of diagram
GRID ?
Returns the definition of all diagrams
GRID:LOG
P1,P2,P3
Defines the diagrams for logic channels
P1=Number of logic channels
P2=Heigh of logic channels
P3=UP or DOWN : position of logic channels
GRID:LOG ?
Returns the definition of logic diagram
GRID:LENGth
P1,P2,P3
GRID:LENG 1,0,100
Defines each diagram
Diagram 1 from 0 to 100mm
P1= number of the diagram
P2= min value ( 0 to max)
max is 250 or 200 (depend of instrument)
P3= max value max ( 0 to max)
GRID:LENGth ?
Returns the definition of diagrams
GRID:CHAnnel
P1,P2,P3
GRID:CHA A4,3,2
Defines the position of a channel
Channel A4 in diagram 3 with a width of
2
P1= Number of the channel
P2= Number of diagram : from 1 to Max
P3= drawing thickness : 1 to 8
GRID:CHAnnel ?
Return the definition of CHANNEL
COLOR
P1,P2,P3
CHAN A2,COLOR 100,100,100
Color of CHANNEL
P1 = red value ( from 0 to 100)
P2=green value
P3= blue value
DEFLOG
P1,P2,P3,P4,P5
Defines logic channels
P1= Number of the logic channel
P2= red value ( from 0 to 100)
P3= green value
P4= blue value
P5= Name of the logic channel
GRID:LOG 50,5,UP;:GRID
2,SEPLOGON
Logic channels are at the top, with height
of 50 mm.
We have 2 diagrams of 100 mm each.
P2=SEPLOGON ou SEPLOGOFF : separated logic channels or not