11

File mode

Page 2

Sample rate: sampling rate of the channels

-

internally rated from 1 s to 20 min

-

externally rated through the logical channel 16

Start: start condition of the acquisition

- Manual: with the F2 key Urge triggering

-

Trigger: with an association of the analogical or logical channels - see Chapter 6
Trigger

- Waiting for: after a delay or at a precise date and time
- Automatic: immediate; automatic stop when the block is full

Stop: stop condition of the acquisition

- Automatic: when the block is full
- Trigger: with an association of the analogical or logical channels - see Chapter 6

Trigger

Trigger mode Double Trigger: see below Memory mode

Following: actions after the end of the acquisition

- Stop: no action
- Ream: waiting for the start condition
- Change Setup: loading a configuration and waiting for the start trigger

11.2.

Limits

Real-time acquisitions on file are limited by the transfer rate between the inputs and the saving
hardware liek an internal hard disk or a USB key.
This limit depends on the type of the acquisition file: binary or ascii.

11.1.1.Binary file

The maximum transfer rate is 1,200,000 bytes/s (1.2 Mbyte/s).

In order to know the number of channels that you can record for a given acquisition period, you
have to calculate:

number of channels = 1,200,000 x acquisition period

example:

acquisition period: 10 s
number of channels = 1,200,000 x 10 s = maximum 12 channels

11.1.2.Ascii file

The maximum transfer rate is 1,000 bytes/s (1.0 Kbyte/s).
The minimum acquisition period is 1ms.

In order to know the number of channels that you can record for a given acquisition period, you
have to calculate:

number of channels = 1 000 x acquisition period