Escort 3016B User Manual
Page 127
16 Technical specifications
Page 8
16.11.4. Security isolation class installation category
Class 1 product
Security
conform with EN61010-1
Pollution degree
2
Installation category (overvoltage category)
Mains input
category II
Measurement input
category III 600 V, overvoltage 6000 V
16.11.5. Electromagnetic compatibility
Assessment criteria for the functioning that were applied for the test:
Criteria A : Normal behaviour within the limits of the specifications
Criteria B : Temporary decrease or self-recovered loss of function
Criteria C : Temporary decrease or loss of function or behaviour that requires the
intervention of the user or a reset of the system
Particular precautions are necessary to preserve the conformity of the product,
particularly the use of shielded cables.
Designation
Accordance test
Specifications
Limits
Criteria
Radiated Emissivity rayonn e
NF EN 55022
30 MHz to 230 MHz
230 MHz to 1 GHz
Measure distance 3 meters
40 dB V/m
47 dB V/m
Class B
Conducted Emissivity
(lines, power supply)
NF EN 55022
0,15 MHz to 0,5 MHz
0,5 MHz to 5 MHz
5 MHz to 30 MHz
66 to 56 dB V QP
56 dB V QP
60 dB V QP
(mean value = QP - 10dB)
Class B
Harmonic Distortion
CEI 61000-3-2
Picture 1 of the standard
Class A
Variation of tensions and flickers
CEI 61000-3-3
chap 5 of the standard
-
CEI 61000-4-2
discharges
Contact
Air
N = +/- 4 kVolts
N = +/- 8 kVolts
Criteria B
CEI 61000-4-3
80MHz to 1 GHz
1,4 GHz to 2 GHz
AM 80% 1000Hz
10 V/m without mod.
Criteria A
CEI 61000-4-8
50 Hz
30 A/m
Criteria B
CEI 61000-4-4
5-50 ns / 5 kHz
Line energy +/- 2 kV
Line Ethernet +/- 1 kV
Line measure input +/- 1 kV
Line earth +/- 1 kV
Criteria B
CEI 61000-4-5
1,2 / 50 s (8/20)
line / line +/- 1 kV
line / earth +/- 2 kV
Criteria B
CEI 61000-4-6
150 kHz to 80 MHz
AM 80% 1000 Hz
Line energy
Line Ethernet
Line input measure
3V
without
mod.
Criteria A
CEI 61000-4-11
Reduction 100%
0,5 cycle with each polarity
Criteria B
0 to 2 kHz
Enveloppe immunity
Access immunity