10 Go/Nogo mode

Page 2

Sample rate: sampling rate of the channels

-

internally rated from 1 s to 20 min

-

externally rated through the logical channel 16

Blocks: division of the internal memory into blocks

-

1, 2, 4, 8, 16, 32, 64 or 128

-

Raz Blocks: deleting all the blocks: the current block is the #1 block

Start: start condition of the acquisition

- Manual: with the F2 key Urge triggering

-

Trigger: with an association of the analogical or logical channels - see Chapter 6
Trigger

- Waiting for: after a delay or at a precise date and time
- Automatic: immediate; automatic stop when the block is full

Pre-triggering: definition of the position of the start trigger in the acquisition

- pre-trigger and post-trigger of the acquisition (before or after Start); see chapter

Memory Mode

Real-time Saving: simultaneous recording of the acquisition into a file

- place and name of the saving file
- maximum length of the file

Impossible is displayed when the saving capabilities are overrun: reduce the

sampling rate or the number of channels

See chapter File mode to know the limitations of the file recording

Following: actions after the end of the acquisition

- Save on file: saving the acquisition into a file if the real-time saving is not validated

(or is impossible)

- Plot: plotting the acquisition
- Stop: no action
- Rearm.: waiting for the start condition
- Change Setup: loading a configuration and waiting for the start trigger

10.2.

Creation of the frame

Make a normal acquisition or display a previously recorded file with the function Memory
output .

In the page Trigger , select the parameter Channels for Go/Nogo then Visualize the
frame
then Modify .

The appliance will then display the current frame. It appears as 2 curves (a min and a max curve)
around the channel selected as a reference for the frame.