Printer function – Fujimoto Fuji Differential Pressure Transmitter FKK 4 User Manual

Page 27

Advertising
background image

21

[Contents of message]

As a result of self-diagnosis, the message below is appeared on the LCD display of HHC,
when there are trouble in the transmitter. For each error, its cause and remedy are suggested.

Printer function

Usable only when a printer is con-
nected.
Refer to the instruction manual of
HHC.

E-2: PRINT OUT

E:

PRINT

< INC >

E-1: PRINT

DATE
TIME

< CL >

E-3: PRINT OUT

PRINT OUT END
< INC >PAPER FEED
< CL >

CL

INC

INC

CL

ENT

INC

ENT

(Indicated when the printer unit is active)

YY : MM : DD
HH : MM

Paper Feed (Printer)

Without Printer Unit

E:

PRINT

< INC >

NO CONNECTION

With Printer Unit

< ENT >

< ENT >

Time (Hour, Minute)

Set Year/Month/Date

F : XMTR EXT.SW

Indication on

digital indicator

CELL FAULT (C1)

CELL FAULT (C9)

(*1)

EEPROM (AMP) FLT

EEPROM (CELL) FLT

TEMP. ALARM

XMTR FAULT

Error of detecting unit

EEPROM error on amplifier side

EEPROM error on cell side

Transmitter temperature is not
within the allowable range (–45
to 90

°C).

Amplifier error

Input pressure is 105% or higher
of setting range.

Input pressure is –1.25% or
lower of setting range.

Replacement of detecting unit

Replacement of amplifier

Replacement of detecting unit

Transmitter temperature is
normalized.

Replacement of amplifier

Properly controlled.

Properly controlled.

Message

Remedy

Cause

FL-1

FL-2

FL-3

T.ALM

FL-1

OVER

Under

ZERO

(*2)

ZERO

(*2)

ZERO

(*2)

(*1) CELL FALUT (C5) to (C9) are appeared in case of “9” in 6th digit code of FKC.