Important – Fujimoto Fuji Differential Pressure Transmitter FKK 4 User Manual

Page 33

Advertising
background image

27

When ENT is pressed at display
of 20 , the following is displayed,

Requirement of setting

1. LP LP2 LP3…LP8 LP9…LP13 LP14 (In the

case that LP1-LP14=All Zero, it is inhibited to be
set enable)

2. If CVa

πCVb, then it must be LPa < LPb. (Note1)

3. If LPa =LPb, then it must be CVa= CVb. (Note 1)

Note 1) a, b show next numeral such as a=1 b=2

or a=2 b=3 or ………a=13 b=14.

or

the following is displayed.

When INC is pressed at display of 3 ,
the following is displayed,

or

When ENT is pressed at display of 4 ,
the following is displayed,

POINT=2 (number of correction) 14
setting err=00 or 01 or 15

When ENT is pressed at display of 15 ,
the following is displayed,

Requirement of setting
-1.25% LP1 < LP2… <
LP14 +110%

When ENT is pressed at display of 17 ,
the following is displayed,

Requirement of setting
-100% CV1 < CV2… <
CV14 +100%

H-2: LINEARIZE

POINT 0

SETTING ERR <CL>

H-3: LINEARIZE

POINT 15
POINT SET

SETTING ERR <CL>

H-5: LINEARIZE

CHANGE

LPM

150.01%

SETTING ERR <CL>

H-5: LINEARIZE

CHANGE

CVM

100.01%

SETTING ERR <CL>

H-3: LINEARIZE
Set LINEARIZE
Point, LP and CV
correctly.

<CL>

H-3: LINEARIZE
Set OUTPUT MODE
LIN-LIN or
SQR-SQR

<CL>

Before performing the LINEARIZE
setting, set either of the following
equations in the OUTPUT mode
(Menu No. 8) and XMTR DISPLAY
(Menu No. G):
OUT = LIN SMTR DISP = LIN or
OUT =SQR XMTR DISP = FLOW

(Note 1)

Note 1)

XMTR DISP = FLOW means the
settings of % FLOW in %display
or of FLOW units in actual-scale
display.

CL

CL

CL

CL

CL

2

Next

parameter

17

INC

INC

INC

18

13

17

18

11

H-4:

LINEARIZE

< INC> < CNG/ENT/CL >

CV1 XXX.XX%

H-4:

LINEARIZE

< INC> < CNG/ENT/CL >

CV3 XXX.XX%

for display

for display

17

INC

11

18

14

17

18

8

H-4:

LINEARIZE

< INC> < CNG/ENT/CL >

CV2 XXX.XX%

H-4:

LINEARIZE

< INC> < CNG/ENT/CL >

CV14 XXX.XX%

CL

ENT

for display

ENT

8

16

15

6

H-5: LINEARIZE

CHANGE

CHANGE

CHANGE

< ENT > <CL >

LPM XXX.XX%

H-5:

LINEARIZE

CHNG OK? < ENT/CL >

LP1-14

for display 7 , 8 , 9 , 10

for display 7 , 8 , 9 , 10

CL

CL

ENT

for display

ENT

8

18

17

H-5: LINEARIZE

CHANGE

CHANGE

CHANGE

< ENT > <CL >

CVM XXX.XX%

H-5:

LINEARIZE

CHNG OK? < ENT/CL >

CV1-14

for display 11, 12 , 13, 14

for display 11, 12 , 13, 14

Important

CV1

CV2

CV3

CV14

INC

DEC

Change of compensated
point for CV

Important

Advertising