Fujimoto Fuji Differential Pressure Transmitter FKK 4 User Manual

Page 31

Advertising
background image

25

When setting of % Flow in %display or
Flow unit in actual scale display, low
flow cut point and low flow cut mode
are displayed ( 22 or 23 ).
When, in the OUTPUT MODE (Menu
No. 8), OUT = SQR is set, already set
low flow cut point and low flow cut
mode are displayed ( 23 ).
With OUT = LIN set, the present low
flow cut point and low flow cut mode
are displayed ( 22 ). Then, enter
<CHANGE>, and the setting can be re-
newed.

1

2

ENT

G-8 : LOW CUT
POINT=XX.XX%FLOW
MODE=YYYYYY
<CHANGE><CL>

22

B

CHNG

1

2

CL

G-9 : POINT CHANGE
XX.XX%FLOW

<ENT><CL>

24

22

CL

ENT

28

CL

ENT

28

CL

or

G-8 : LOW CUT
POINT=XX.XX%FLOW
MODE=YYYYYY
<CL>

23

1

2

CL

or

OUT=LIN at
Menu No.8

OUT=SQR at Menu No.8

G-C : MODE CHANGE
<1>LINEAR
<2>ZERO
<1><2><CL>

28

27

CL

ENT

G-9 : POINT CHANGE
XX.XX%FLOW

CHNG OK?<ENT/CL>

25

G-A : POINT WRITE
XX.XX%FLOW

26

24

CL

G-B : LOW CUT MODE
LINEAR

<CHANGE><CL>

27

CHNG

CHNG

1

2

CL

or

G-B : LOW CUT MODE
ZERO

<CHANGE><CL>

27

G-D : MODE CHANGE
LINEAR

CHNG OK?<ENT/CL>

29

G-B : MODE CHANGE
ZERO

CHNG OK?<ENT/CL>

29

G-D : MODE WRITE
LINEAR

30

G-B : MODE WRITE
ZERO

30

1

2

CL

or

Note2) CUT POINT

and
CUT MODE
are set at Menu No.8

YYYYYY=LINEAR or ZERO

Advertising