background image

43

Pin plug

Remove pin
plug fitted to
indicator.

Attachment

Mount the supplied
attachment on
indicator.

Studs

Indicator attached
to transmission
unit side.

Remove the transmission
unit side cover with
window and indicator.

Remove the terminal
unit side cover.

Mount the studs
supplied with the
attachment on
the terminal
block.

Connect the red wire (+)
of the indicator to CK+
and the black wire (-) of
the indicator to CK-, then
attach the cover with
window.

Indicator attached to
terminal board side.

Attach the cover.

Change of field indicator position

It is sometimes preferable to mount the indicator on the terminal block side due to
installation location.
In such a case, the following mounting procedure should be followed for analog indicator.

CAUTION

Avoid the following procedure in an explosionproof area.

Digital indicator cannot be mounted on the terminal block side.