Change of transmission unit position caution, Caution – Fujimoto Fuji Differential Pressure Transmitter FKK 4 User Manual

Page 50

Advertising
background image

44

Wiring is sometimes difficult depending on the installation loca-
tion. In such a case, it is convenient to carry out the following.
Before turning the transmission unit, remove the electronics
unit.
The transmission unit is secured by 2 hexagonal socket bolts.
Loosen the bolts, turn the transmission unit at 90˚ or 180˚ in the
clockwise or counterclockwise direction and fix it by the screws.
Then, carry out wiring.

Hex.
socket bolt

Turn at 90°

in the clockwise.

Turn at 90°

in the counter-
clockwise.

In case of an analog indicator, it can be turned

±180˚ in 90˚

increments because it is connected with a pin plug.

Digital indicator does not allow change of its mounting
direction.

0

20

40

60

80

100

0

20

40

60

80

100

0

20

40

60

80

100

0

20

40

60

80

100

Change of transmission unit position

CAUTION

Avoid the following procedure in an explosionproof area.

Important

If the transmission unit has been turned
excessively without removing the
electronics unit, straighten the ribbon
cable which connects the electronics unit
in the transmission unit and the detecting
unit, and set the transmission unit again.

Change of indicator angle

CAUTION

Avoid the following procedure in an
explosionproof area.

Advertising