Velleman HPS50 User Manual

Page 32

Advertising
background image

32

HPS50 PersonalScope

TM

6. dBV measurement (dBV ac).
The measured signal (ac only) is converted to dBv
(0dB= 1V).

7. dBm measurement (dBm ac).
The measured signal (ac only) is converted to dBm
(0dB= 0.775V).

8. dB measurement (dB ac).
The measured signal (ac only) is converted to dB
(0dB= dBref*)

9. True RMS readout (Vrms ac+dc)
The true RMS value of the wave (ac+dc) is
calculated and converted to voltage.

10. dBV measurement (dBV ac+dc).
The measured signal (ac+dc) is converted to dBv
(0dB= 1V).

11. dBm measurement (dBm ac+dc).
The measured signal (ac+dc) is converted to dBm
(0dB= 0.775V).

12. dB measurement (dB ac+dc).
The measured signal (ac+dc) is converted to dB
(0dB=dBref*)

Fig 26

9

10

8

7

12

11

Meter readout

*dB ref
Select dBref to set the user defined reference for
dB measurement, the selected meter will be set for
dB measurement