HPS50 PersonalScope

TM

Trigger function

Fig 42

Press the „trigger“-key.

Change the triggermode (norm, run, once or
roll).

Use the left cursorkey to toggle the trigger slope

Use the up/down cursorkeys to shift the vertical
trigger position (1).

Use the right arrow key to force a manual trig-
gering.

trig

1

THE TRIGGER SETUP

Note:

Changing the timebase or sensitivity switches
the autorange mode immediately off. (Fig 42)

The timebase or sensitivity cannot be changed
into hold mode.

Pressing the ‘t-V/div’ key into hold mode
toggles the screen between the two stored
waveforms.

At higher timebase (1µs and faster) the scope
uses oversampling mode, only repetitive
signals are correctly displayed.

Use the minimum time base (250ns) as a
starting point when measuring a signal and
select longer time bases until the signal is
displayed properly. Otherwise the display
may not correctly reflect the signal under
measurement due to aliasing.