YSI 556 MPS User Manual

Page 28

Advertising
background image

Probe Module

Section 3

YSI 556 MPS

YSI Incorporated

Page 22

O-Rings

Sensor nut

Figure 3.4 O-ring Lubrication

CAUTION: Make sure that there are NO contaminants between the o-ring

and the sensor. Contaminants that are present under the o-ring may cause
the o-ring to leak.

5. Be sure the probe module sensor port is free of moisture and

then insert the sensor into the correct port. Gently rotate the
sensor until the two connectors align.

6. With the connectors aligned, screw down the sensor nut using

the sensor installation tool.

Figure 3.5

Sensor Installation

CAUTION: Do not cross thread the sensor nut. Tighten the nut until it is

flush with the face of the probe module bulkhead. Do not over tighten.

Figure 3.6

Bulkhead Seating

7. Repeat steps 3-6 for any other sensors.