YSI 556 MPS User Manual

Page 50

Advertising
background image

Section 6

Calibrate

YSI 556 MPS

YSI Incorporated

Page 44

Figure 6.10 DO mg/L Entry Screen

4. Place the probe module in water with a known DO

concentration.

NOTE: Be sure to completely immerse all the sensors.

5. Use the keypad to enter the known DO concentration of the

water.

6. Press Enter. The Dissolved Oxygen mg/L Calibration Screen

is displayed.

Figure 6.11 DO mg/L Calibration Screen