Section 6

Calibrate

YSI 556 MPS

YSI Incorporated

Page 40

Figure 6.6 Calibrated

14. Press Enter. This returns you to the Conductivity Calibrate

Selection Screen, See Figure 6.3 Conductivity Calibration
Selection Screen..

15. Press Escape to return to the calibrate menu. See Figure 6.2

Calibrate Screen .

16. Rinse the probe module and sensors in tap or purified water

and dry.

6.2.3 Dissolved Oxygen Calibration

This procedure calibrates dissolved oxygen. Calibrating any one option (%
or mg/L) automatically calibrates the other.

1. Go to the calibrate screen as described in Section 6.2.1

Accessing the Calibrate Screen.

NOTE: The instrument must be on for at least 10 - 15
minutes to polarize the DO sensor before calibrating.

2. Use the arrow keys to highlight the Dissolved Oxygen

selection. See Figure 6.2 Calibrate Screen.

3. Press Enter. The dissolved oxygen calibration screen is

displayed.