9.8 Adding Data to Existing Files

In order to add new data to an existing file, the current logging and sensor
setup must be exactly the same as when the file was created. The following
settings must be the same:

Sensors enabled (refer to Section 4 Sensors)

Store Barometer (refer to Section 9.3 Storing Barometer
Readings
)

Store Site Number (refer to Section 9.4 Creating a Site
List
)

If the current logging setup is not exactly the same as when the file was
created, a parameter mismatch screen is displayed.

Figure 9.14 Parameter Mismatch Screen

NOTE: The right column shows parameters used when the
file was created. The left column shows current parameters.

1. Press the Down Arrow key to scroll down and find the

mismatch(es).

2. Use the following chart to resolve the mismatch(es).