Real-time data – YSI 556 MPS User Manual
Page 60
YSI 556 MPS YSI Incorporated
Page 54
7. Run
The Run screen displays data from the sensors in real-time and allows the
user to log sample data to memory for later analysis. Refer to Section 9
Logging for details on logging sample data.
7.1 Real-Time Data
NOTE: Before measuring samples you must prepare the probe module
(refer to Section 3.4 Preparing the Probe Module), attach the probe module
to the instrument (refer to Section 3.6 Instrument/Cable Connection) and
calibrate the sensors (refer to Section 6 Calibrate).
1. Press the On/off key.
OR select Run from the main menu to display the run screen.
Figure 7.1 Run Screen
2.
Make sure the probe sensor guard is installed.
3. Place the probe module in the sample. Be sure to completely
immerse all the sensors.
4. Rapidly move the probe module through the sample to
provide fresh sample to the DO sensor.
5. Watch the readings on the display until they are stable.
6. Refer to Section 9 Logging for instructions on logging sample
data.