Maintenance

Section

11

YSI Incorporated

YSI 556 MPS

Page 93

Probe Sensor
Guard

Set Screw

DO Sensor

Hex Key
Wrench

F

igure 11.1 DO Sensor Replacement

4. Insert the new DO sensor module. Make sure that the inside

of the probe module sensor port and the o-ring on the sensor
are clean, with no contaminants, such as grease, dirt, or hair.
The DO sensor is keyed, or has a flat side, so that it cannot be
aligned improperly.

NOTE: Make sure the DO sensor bottoms out before the set
screw is inserted.

5. Insert the set screw into the small hole in the side of the probe

module bulkhead, and turn clockwise to rethread.

CAUTION: Make sure that you do not cross-thread the set

screw. Use the hex key wrench to tighten the screw in properly,
making sure that the screw does not stick out of the side of the
probe module bulkhead. The probe sensor guard will not thread
on properly and damage may result if the screw is allowed to
stick out.