YSI 556 MPS User Manual

Page 98

Advertising
background image

Section 11

Maintenance

YSI 556 MPS

YSI Incorporated

Page 92

Gold Cathode Cleaning

For correct sensor operation, the gold cathode must be textured properly. It
can become tarnished or plated with silver after extended use. The gold
cathode can be cleaned by using the adhesive backed sanding disc and tool
provided in the YSI 5238 Probe Reconditioning Kit.

Using the sanding paper provided in the YSI 5238 Probe Reconditioning
Kit, wet sand the gold with a twisting motion about 3 times or until all silver
deposits are removed and the gold appears to have a matte finish. Rinse the
cathode with clean water after sanding, followed by wiping thoroughly with
a wet paper towel. If the cathode remains tarnished, contact your dealer or
YSI Customer Service. Refer to Appendix E Customer Service.

NOTE: After cleaning, a new membrane cap must be installed. Refer to
Section 3.4.3 Membrane Cap Installation.

11.1.2 DO Sensor Replacement

1. Remove the probe sensor guard.

CAUTION: Thoroughly dry the sensor so that no water enters the probe

module sensor port when the sensor is removed.

2. Insert the long end of the hex key wrench into the small hole

in the side of the probe module bulkhead. Turn the wrench
counterclockwise and remove the screw. (You do not have to
remove the screw all the way to release the sensor.)

3. Pull the old DO sensor module straight out of the probe

module body.

NOTE: The DO sensor is not threaded, it is keyed, so it
cannot be removed by twisting.