Section 6
Calibrate
YSI 556 MPS
YSI Incorporated
Page 36
Figure 6.2 Calibrate Screen
6.2.2 Conductivity Calibration
This procedure calibrates specific conductance (recommended),
conductivity and salinity. Calibrating any one option automatically
calibrates the other two.
1. Go to the calibrate screen as described in Section
6.2.1Accessing the Calibrate Screen..
2. Use the arrow keys to highlight the Conductivity selection.
See Figure 6.2 Calibrate Screen.
3. Press Enter. The Conductivity Calibration Screen is
displayed.