Page 39

10. Use the keypad to enter the calibration value of the

standard you are using.

NOTE: Be sure to enter the value in mS/cm at 25

°C.

11. Press Enter. The Conductivity Calibration Screen is

displayed.

Figure 6.5 Conductivity Calibration Screen

12. Allow at least one minute for temperature equilibration before

proceeding. The current values of all enabled sensors will
appear on the screen and will change with time as they
stabilize.

13. Observe the reading under Specific Conductance. When the

reading shows no significant change for approximately 30
seconds, press Enter. The screen will indicate that the
calibration has been accepted and prompt you to press Enter
again to Continue.