YSI 556 MPS User Manual

Page 43

Advertising
background image

Calibrate

Section

6

YSI Incorporated

YSI 556 MPS

Page 37

Figure 6.3 Conductivity Calibration Selection Screen

4. Use the arrow keys to highlight the Specific Conductance

selection.

5. Press Enter. The Conductivity Calibration Entry Screen is

displayed.

Figure 6.4 Conductivity Calibration Selection Screen

6. Place the correct amount of conductivity standard (see Table

6.1 Calibration Volumes) into a clean, dry or pre-rinsed
transport/calibration cup.