YSI 556 MPS User Manual

Page 47

Advertising
background image

Calibrate

Section

6

YSI Incorporated

YSI 556 MPS

Page 41

Figure 6.7 DO Calibration Screen

DO

Calibration

in

%

Saturation

1. Use the arrow keys to highlight the DO% selection.

2. Press Enter. The DO Barometric Pressure Entry Screen is

displayed.

Figure 6.8 DO Barometric Pressure Entry Screen

3. Place approximately 3 mm (1/8 inch) of water in the bottom

of the transport/calibration cup.

4. Place the probe module into the transport/calibration cup.