Return to factory settings – YSI 556 MPS User Manual

Page 58

Advertising
background image

Section 6

Calibrate

YSI 556 MPS

YSI Incorporated

Page 52

11. Observe the reading under ORP, when the reading shows no

significant change for approximately 30 seconds, press Enter.
The screen will indicate that the calibration has been accepted
and prompt you to press Enter again to Continue.

12. Press Enter. This returns you to the Calibrate Screen. See

Figure 6.2 Calibrate Screen.

13. Rinse the probe module and sensors in tap or purified water

and dry.

6.3 Return to Factory Settings.

1. Go to the calibrate screen as described in Section 6.2.1

Accessing the Calibrate Screen.

2. Use the arrow keys to highlight the Conductivity selection.

See Figure 6.2 Calibrate Screen.

NOTE: We will use the Conductivity sensor as an example;
however, this process will work for any sensor.

3. Press Enter. The Conductivity Calibration Selection Screen is

displayed. See Figure 6.3 Conductivity Calibration Selection
Screen.

4. Use the arrow keys to highlight the Specific Conductance

selection.

5. Press Enter. The Conductivity Calibration Entry Screen is

displayed. See Figure 6.4 Conductivity Calibration Entry
Screen.

6. Press and hold the Enter key down and press the Escape key.