8. Use the keypad to enter the correct value of the calibration

solution you are using at the current temperature. Refer to
Table 6.2 Zobell Solution Values.

Temperature °C

Zobell Solution Value, mV

-5 270.0

0 263.5
5 257.0

10 250.5
15 244.0
20 237.5
25 231.0
30 224.5
35 218.0
40 211.5
45 205.0
50 198.5

Table 6.2 Zobell Solution Values

9. Press Enter. The ORP calibration screen is displayed.

Figure 6.16 DO Cal Screen

10. Allow at least one minute for temperature equilibration before

proceeding. The current values of all enabled sensors will
appear on the screen and will change with time as they
stabilize.

NOTE: Verify that the temperature reading matches the value
you used in Table 6.2 Zobell Solution Values.