YSI 556 MPS User Manual

Page 48

Advertising
background image

Section 6

Calibrate

YSI 556 MPS

YSI Incorporated

Page 42

NOTE: Make sure that the DO and temperature sensors are
not immersed in the water.

5. Engage only 1 or 2 threads of the transport/calibration cup to

ensure the DO sensor is vented to the atmosphere.

6. Use the keypad to enter the current local barometric pressure.

NOTE: If the unit has the optional barometer, no entry is
required.

NOTE: Barometer readings that appear in meteorological
reports are generally corrected to sea level and must be
uncorrected before use (refer to Section 10.10 Calibrate
Barometer, Step 2
).

7. Press Enter. The DO% saturation calibration screen is

displayed.

Figure 6.9 DO Sat Calibration Screen

8. Allow approximately ten minutes for the air in the

transport/calibration cup to become water saturated and for
the temperature to equilibrate before proceeding.