4.2.2

Measuring the pH value

Setting the resolution

AutoRead AR

(drift control)

The AutoRead function (drift control) checks the stability of the mea-
surement signal. The stability has a considerable impact on the repro-
ducibility of the measured values.

For identical measurement conditions, the following criteria apply:

1

Perform the preparatory activities according to section 4.2.1.

2

Immerse the pH electrode in the test sample.

3

If required, enter the temperature of the test sample with <

>

<

> while keeping the <RUN/ENTER> key depressed.

4

Press the <

> <> keys until pH appears in the status display.

The pH value appears on the display.

mV/pH

mol/L

8

Auto Store

TempError

RCL

LoBat

AutoCal DIN

CalError

ARng AR

AutoCal TEC ConCal

Time

Baud

Day.Month No.

Ident

Year

1

U

%

°F

S

mg/L

% ppm

ISE

8

4

2

2

99

6

°C

TP

pH

1

Press the <RUN/ENTER> key and hold it down.

2

Press the <M> key. The measured values are displayed with a
high resolution, e.g. pH = 4.012.

3

Press the <RUN/ENTER> and <M> key once more.
The measured values are displayed with a low resolution, e.g.
pH = 4.01.

Reproducibility

Response time

Better than 0.02

> 30 seconds

1

Call up the pH measuring mode with <M> and <

> <> .

2

Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).