4 clearing the memory, Clearing the memory, Rc l – Xylem pH/ION 3400i ION SELECTIVE ELECTRODE ANALYZER User Manual
Page 52: F °c

Operation
pH/ 3400i
52
4.6.4
Clearing the memory
With this function, you can delete stored data records. 500 storage lo-
cations will then be available again.
Note
The Clear memory function only appears when there are data records
stored in the memory. Otherwise, the measuring instrument automati-
cally switches to the last active measuring mode.
Proceed as follows to clear all data records:
Note
The calibration data remain stored and can be called up.
1
Switch off the measuring instrument.
2
Press the <STO> key and hold it down.
3
Press the <ON/OFF> key.
The display test appears briefly on the display.
4
Confirm the clearing process with <RUN/ENTER>.
Pressing any other key prevents the clearing, the data records
will remain stored.