2 switching on autostore (int 1), Switching on autostore (int 1), 1nt i – Xylem pH/ION 3400i ION SELECTIVE ELECTRODE ANALYZER User Manual
Page 44: F °c

Operation
pH/ 3400i
44
4.6.2
Switching on AutoStore (Int 1)
The save interval (Int 1) determines the chronological interval between
automatic save processes. After the fixed interval has expired, the cur-
rent data record is transmitted to the internal storage and to the inter-
face.
Setting the
save interval
The default setting for the save interval (Int 1) is OFF.
By this, the AutoStore function is switched off.
To switch the function on, set an interval (5 s, 10 s, 30 s, 1 min, 5 min,
10 min, 15 min, 30 min, 60 min):
1
Press the <RUN/ENTER> key and hold it down.
2
Press the <STO> key. Int 1 appears on the display.
3
Set the required interval between the saving procedures with
<
▲> <▼> (Selection: 5 s, 10 s, 30 s, 1 min, 5 min, 10 min,
15 min, 30 min, 60 min).
4
Confirm with <RUN/ENTER>.
The number of free memory locations appears on the display.
5
Confirm with <RUN/ENTER>.
The prompt for the ID number appears on the display.
No.
LoBat
Day.Month
Year
Ident
UpH
RCL
AR
ARng
8
%
°F
°C
TP
mV/pH
mol/L
TempError
AutoCal DIN
CalError
AutoCal TEC ConCal
Time
Baud
1
S
mg/L
% ppm
ISE
1
nt
I
f
Store
O f
Auto
U
I
No.
LoBat
Day.Month
Year
Ident
pH
RCL
AR
ARng
%
°F
°C
TP
mV/pH
mol/L
TempError
AutoCal DIN
CalError
AutoCal TEC ConCal
Time
Baud
1
S
mg/L
% ppm
ISE
9
49
E
Store
r E
Auto
F