2 switching on autostore (int 1), Switching on autostore (int 1), 1nt i – Xylem pH/ION 3400i ION SELECTIVE ELECTRODE ANALYZER User Manual

Page 44: F °c

Advertising
background image

Operation

pH/ 3400i

44

4.6.2

Switching on AutoStore (Int 1)

The save interval (Int 1) determines the chronological interval between
automatic save processes. After the fixed interval has expired, the cur-
rent data record is transmitted to the internal storage and to the inter-
face.

Setting the

save interval

The default setting for the save interval (Int 1) is OFF.
By this, the AutoStore function is switched off.
To switch the function on, set an interval (5 s, 10 s, 30 s, 1 min, 5 min,
10 min, 15 min, 30 min, 60 min):

1

Press the <RUN/ENTER> key and hold it down.

2

Press the <STO> key. Int 1 appears on the display.

3

Set the required interval between the saving procedures with
<

> <> (Selection: 5 s, 10 s, 30 s, 1 min, 5 min, 10 min,

15 min, 30 min, 60 min).

4

Confirm with <RUN/ENTER>.
The number of free memory locations appears on the display.

5

Confirm with <RUN/ENTER>.
The prompt for the ID number appears on the display.

No.

LoBat

Day.Month

Year

Ident

UpH

RCL

AR

ARng

8

%

°F

°C

TP

mV/pH

mol/L

TempError

AutoCal DIN

CalError

AutoCal TEC ConCal

Time

Baud

1

S

mg/L

% ppm

ISE

1

nt

I

f

Store

O f

Auto

U

I

No.

LoBat

Day.Month

Year

Ident

pH

RCL

AR

ARng

%

°F

°C

TP

mV/pH

mol/L

TempError

AutoCal DIN

CalError

AutoCal TEC ConCal

Time

Baud

1

S

mg/L

% ppm

ISE

9

49

E

Store

r E

Auto

F

Advertising