7 technical data, Technical data, Chapter 7 t – Xylem pH/ION 3400i ION SELECTIVE ELECTRODE ANALYZER User Manual
Page 71: Echnical, Data, Ee chapter 7 t, 7technical data
Advertising

pH/ 3400i
Technical data
71
7
Technical data
Dimensions and weight
Mechanical structure
Electrical safety
Ambient
conditions
pH/ORP measuring
ranges
Precision of pH/ORP
(± 1 digit)
Length [mm]
172
Width [mm]
80
Height [mm]
37
Weight [kg]
Approx. 0.3
Type of protection
IP 66
Protective class
III
Storage
- 25 °C ... + 65 °C
Operation
-10 °C ... + 55 °C
Climatic class
2
Measuring range
Resolution
pH
- 2.000 ... + 19.999
- 2.00 ... + 19.99
0.001
0.01
U [mV]
- 999.9 ... + 999.9
- 1999 ... + 1999
0.1
1
T [°C]
- 5.0 ... + 105.0
0.1
pH (± 2 pH units from
the calibration point)
± 0.003 at + 15 °C ... + 35 °C
U [mV]
± 0.2 at + 15 °C ... + 35 °C
± 0.1
T [°C]
NTC 30:
± 0.1
PT 1000:
± 0.5 at 0 °C ... 15 °C
± 0.1 at 15 °C ... 35 °C
± 1 at 35 °C ... 55 °C
Advertising