Xylem pH/ION 3400i ION SELECTIVE ELECTRODE ANALYZER User Manual
Page 35

pH/ 3400i
Operation
35
TP display
If you use a temperature sensor, a flashing TP display indicates that the
difference in temperature of the standard solutions is too high (more
than 2 °C).
Displaying the
calibration data
You can view the data of the last calibration on the display. The pro-
ceeding is described on page 48.
Printing the
calibration protocol
The calibration protocol contains the calibration data of the current cal-
ibration. You can transmit the calibration protocol to a printer via the se-
rial interface (see page 51).
Note
You can automatically print a calibration protocol after the calibration.
To do so, connect a printer to the interface according to section 4.7.2
before calibrating. After a valid calibration, the record is printed.
Sample printout:
Calibration evaluation
After the calibration, the measuring instrument evaluates the calibra-
tion. The slope (in mV) is evaluated. The probe symbol shows the cal-
ibration evaluation.
CALIBRATION PROTOCOL
02.03.99 / 16:20
Device No.: 1234
CALIBRATION ISE
Cal time: 01.03.99 / 15:10
Std 1 10,00 mg/l
Std 2 20,00 mg/l
Std 3 50,00 mg/l
C1 0,5 mV 25.0°C
C2 8,4 mV 24.9°C
C3 21,2 mV 25.0°C
S1 26.4 mV
S2 32,1 mV
Probe: +++
Display
Slope [mV]
50.0 ... 70.0
25.0 ... 35.0
E3
Eliminate the error according to
chapter 6 W
HAT
OT
DO
IF
...