2 autocal din, Autocal din, 1c d – Xylem pH/ION 3400i ION SELECTIVE ELECTRODE ANALYZER User Manual

Page 26

Advertising
background image

Operation

pH/ 3400i

26

4.3.2

AutoCal DIN

For this procedure, use two different standard buffer solutions accord-
ing to DIN 19266 (type A, C, D or F with pH values at 25 °C of: 1.679 /
4.006 / 6.865 / 9.180).

Note
Skip the steps 2 and 7 if you use a pH electrode with a temperature
sensor.

1

Press the <CAL> key repeatedly until the display Cd1 and the
function display AutoCal DIN appear. The sensor symbol dis-
plays the evaluation of the last calibration (or no sensor symbol
in the delivery state or after the measurement parameter has
been reset).

2

If required, enter the temperature of the first buffer solution with
<

> <> while keeping the <RUN/ENTER> key depressed.

3

Immerse the pH electrode in the first buffer solution.

4

Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value ap-
pears on the display. Example:

AutoCal TEC

6

mV/pH

mol/L

8

Auto Store

RCL

LoBat

CalError

ARng AR

ConCal

Time

Baud

Day.Month No.

Ident

Year

1

U

%

°F

S

TempError

mg/L

% ppm

ISE

8

4

2

1

C d

°C

TP

pH

AutoCal DIN

U

AutoCal TEC

AutoCal TEC

mV

U

mV/pH

mol/L

8

Auto Store

TempError

RCL

LoBat

CalError

ARng

ConCal

Time

Baud

Day.Month No.

Ident

Year

1

%

°F

S

mg/L

% ppm

ISE

8

4

2

°C

TP

mV/pH

mol/L

8

Auto Store

TempError

RCL

LoBat

CalError

ARng

ConCal

Time

Baud

Day.Month No.

Ident

Year

1

%

°F

S

mg/L

% ppm

ISE

8

4

2

6

0

°C

TP

7 7

7

pH

mV

AR

0

AR

AutoCal DIN

AutoCal DIN

pH

1

4

Advertising