5 calibration interval (int 3), Calibration interval (int 3) – Xylem pH/ION 3400i ION SELECTIVE ELECTRODE ANALYZER User Manual

Page 41

Advertising
background image

pH/ 3400i

Operation

41

4.5

Calibration interval (Int 3)

The flashing probe symbol reminds you to calibrate regularly (pH/
ORP). When the fixed calibration interval (Int 3) has expired, the probe
symbol flashes. It is still possible to measure. The calibration interval is
set to 7 days in the factory.

Note
To ensure the high measuring accuracy of the measuring system, cal-
ibrate after the calibration interval has expired.

Setting the calibration

interval

You can change the interval set in the factory (1 ... 999 days):

1

Switch off the measuring instrument.

2

Press the <M> key and hold it down.

3

Press the <ON/OFF> key.
The display test appears briefly on the display. After this, the
measuring instrument automatically switches over to configu-
ration.

4

Press the <RUN/ENTER> key repeatedly until Int 3 appears.
Example:

5

Set the required interval (in days) until the next calibration with
<

> <>.

6

Confirm with <RUN/ENTER>.

7

Switch to the measuring mode with <M>.