What to do if...

65

6

What to do if...

6.1

pH system messages

Error message

0FL

0FL

0FL

0FL

Error message

E3

E3

E3

E3

Cause

Remedy

pH electrode:

– Not connected

– Connect electrode

– Air bubble in front of the

diaphragm

– Remove air bubble

– Air in the diaphragm

– Extract air or

moisten diaphragm

– Cable broken

– Replace electrode

– Gel electrolyte dried out

– Replace electrode

Cause

Remedy

Electrode

– Diaphragm contaminated

– Clean diaphragm

– Membrane contaminated

– Clean membrane

– Moisture in the plug

– Dry plug

– Electrolyte out of date

– Replenish electrolyte or

replace electrode

– Electrode worn out

– Replace electrode

– Electrode broken

– Replace electrode

Measuring instrument:

– Incorrect calibration procedure

– Select correct procedure

– Incorrect solution temperature

(without temperature sensor)

– Set up correct temperature

– Socket damp

– Dry socket