2 measuring the ion concentration, Measuring the ion concentration – Xylem pH/ION 3400i ION SELECTIVE ELECTRODE ANALYZER User Manual

Page 33

Advertising
background image

pH/ 3400i

Operation

33

The temperature measurement is absolutely essential for a reproduc-
ible ISE measurement. If the measurement is made without a temper-
ature sensor, proceed as follows:

Note
For ISE measurements and ISE calibration, it is not possible to enter
temperature values manually.

4.4.2

Measuring the ion concentration

To perform ion concentration measurements, proceed as follows:

AutoRead AR

(Drift control)

The AutoRead function (drift control) checks the stability of the mea-
surement signal. The stability has a considerable effect on the repro-
ducibility of the measured value.

For identical measurement conditions, the following criteria apply:

1

Measure the current temperature using a thermometer.

2

Bring the test sample and calibration standards to the same
temperature (± 2 °C).

1

Perform the preparatory activities according to section 4.4.1.

2

Immerse the ISE electrode and, if necessary, the temperature
sensor in the test sample.

UpH

mV/pH

mol/L

8

Auto Store

TempError

RCL

LoBat

AutoCal DIN

CalError

AR

AutoCal TEC

ConCal

Time

Baud

Day.Month No.

Ident

Year

1

%

°F

S

% ppm

8

0

2

7

15

0

°C

ISE

mg/L

ARng

TP

Reproducibility

Response time

better ± 0.05 mV

> 30 seconds