2c d – Xylem pH/ION 3400i ION SELECTIVE ELECTRODE ANALYZER User Manual

Page 27

Advertising
background image

pH/ 3400i

Operation

27

Note
At this point, the AutoCal DIN calibration can be terminated with <M>.
This corresponds to a single-point calibration. When doing so, the in-
strument uses the Nernst slope (-59,16 mV/pH at 25 °C) and deter-
mines the asymmetry of the electrode.

5

When the measured value is stable, Cd2 appears.

6

Thoroughly rinse the electrode with deionized water.

7

If required, enter the temperature of the second buffer solution
with <

> <> while keeping the <RUN/ENTER> key de-

pressed.

8

Immerse the pH electrode in the second buffer solution.

9

Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value ap-
pears on the display. Example:

AutoCal TEC

6

mV/pH

mol/L

8

Auto Store

RCL

LoBat

CalError

ARng AR

ConCal

Time

Baud

Day.Month No.

Ident

Year

1

U

%

°F

S

TempError

mg/L

% ppm

ISE

8

4

2

2

C d

°C

TP

pH

AutoCal DIN

ISE

U

AutoCal TEC

AutoCal TEC

mV

U

mV/pH

mol/L

8

Auto Store

TempError

RCL

LoBat

CalError

ARng

ConCal

Time

Baud

Day.Month No.

Ident

Year

1

%

°F

S

mg/L

% ppm

8

4

2

°C

TP

mV/pH

mol/L

8

Auto Store

TempError

RCL

LoBat

CalError

ARng

ConCal

Time

Baud

Day.Month No.

Ident

Year

1

%

°F

S

mg/L

% ppm

ISE

8

4

2

0

8

°C

TP

2 3

7

pH

mV

AR

1

AR

AutoCal DIN

AutoCal DIN

pH

1

9

Advertising