Operation

37

Two-point calibration

1

Press the <CAL> key. The previously set up concentration of
the first standard solution and ConCal appear on the display.

2

Enter the current concentration of the first standard solution
with <

> <>.

3

Immerse the ISE electrode and, if necessary, the temperature
sensor in the first standard solution.

4

Press the <RUN/ENTER> key.
The electrode voltage appears on the display, the AR display
indicator flashes.

5

As soon as a stable value is achieved, AR stops flashing. The
previously set up concentration of the second standard solution
appears on the display.