Two-point calibration
1
Press the <CAL> key. The previously set up concentration of
the first standard solution and ConCal appear on the display.
2
Enter the current concentration of the first standard solution
with <
▲> <▼>.
3
Immerse the ISE electrode and, if necessary, the temperature
sensor in the first standard solution.
4
Press the <RUN/ENTER> key.
The electrode voltage appears on the display, the AR display
indicator flashes.
5
As soon as a stable value is achieved, AR stops flashing. The
previously set up concentration of the second standard solution
appears on the display.