2d t – Xylem pH/ION 3400i ION SELECTIVE ELECTRODE ANALYZER User Manual

Page 38

Advertising
background image

Operation

pH/ 3400i

38

6

Enter the current concentration of the second standard solution
with <

> <>.

7

Thoroughly rinse the ion selective electrode and, if necessary,
the temperature sensor with deionized water.

8

Immerse the ion selective electrode and, if necessary, the tem-
perature sensor in the second standard solution.

9

Press the <RUN/ENTER> key.
The electrode voltage appears on the display, the AR display
indicator flashes.

10

As soon as a stable value is reached, the AR display indicator
stops flashing. The slope (mV) of the electrode after the two-
point calibration appears on the display.
The probe symbol shows the evaluation of the two-point cali-
bration for the measuring range I.