2d t – Xylem pH/ION 3400i ION SELECTIVE ELECTRODE ANALYZER User Manual
Page 38

Operation
pH/ 3400i
38
6
Enter the current concentration of the second standard solution
with <
▲> <▼>.
7
Thoroughly rinse the ion selective electrode and, if necessary,
the temperature sensor with deionized water.
8
Immerse the ion selective electrode and, if necessary, the tem-
perature sensor in the second standard solution.
9
Press the <RUN/ENTER> key.
The electrode voltage appears on the display, the AR display
indicator flashes.
10
As soon as a stable value is reached, the AR display indicator
stops flashing. The slope (mV) of the electrode after the two-
point calibration appears on the display.
The probe symbol shows the evaluation of the two-point cali-
bration for the measuring range I.