2 ise system messages, Ise system messages – Xylem pH/ION 3400i ION SELECTIVE ELECTRODE ANALYZER User Manual
Page 67

pH/ 3400i
What to do if...
67
6.2
ISE system messages
Error message
0FL
0FL
0FL
0FL
Error message
E3
E3
E3
E3
Concentration
display flashes
Cause
Remedy
Ion-selective electrode:
– Not connected
– Connect the electrode
– Cable broken
– Replace the electrode
Cause
Remedy
Ion-selective electrode:
– Slope not in the range
50 ... 70 mV or
23 ... 35 mV
– Replace the electrode
– recalibrate
– Moisture in the plug
– Dry plug
– Electrode obsolete
– Replace the electrode
– Electrode broken
– Replace the electrode
Measuring instrument:
– Calibration procedures:
Incorrect order of the
standards with 3-point
calibration
– Select the correct order
– Socket damp
– Dry socket
Cause
Remedy
In measurement:
– The measured value is outside
the range determined through
calibration
– Select the calibration
standards so that the
measured value is in between
them
In calibration:
– Calibration standards lie too
closely together. Select other
calibration standards
U2
U1
–
5mV
≤
U3
U2
–
5mV
≤