Transmitting data
You have the following possibilities of transmitting data:
z One of the following options:
– With the AutoStore function (page 44), measured values are
periodically saved internally (save interval Int 1) and output on the
interface.
– With the Data transmission interval function (Int 2), measured
values are periodically output on the interface (see below).
z With the Output data storage function (page 46), calibration data or
saved measured values are output on the interface.
4.7.1
Data transmission interval (Int 2)
The interval to the data transmission (Int 2) determines the chronolog-
ical interval between automatic data transmissions. After the selected
interval expires, the current data record is transmitted to the interface.
Note
The setting of the interval (Int 2) is only effective if the save interval
(AutoStore function) is switched off!
Setting the
Data transmission
interval
The default setting for the interval is OFF.
To start the data transmission, set an interval (5 s, 10 s, 30 s, 1 min, 5
min, 10 min, 15 min, 30 min, 60 min):
1
Press the <RUN/ENTER> key and hold it down.
2
Press the <RCL> key. Int 2 appears on the display.
3
Set the required interval between the saving procedures with
<
▲> <▼>.
4
Confirm with <RUN/ENTER>.
The measuring instrument switches to the last active measur-
ing mode.