6 saving, 1 saving manually, Saving 4.6.1 – Xylem pH/ION 3400i ION SELECTIVE ELECTRODE ANALYZER User Manual
Page 42: Saving manually

Operation
pH/ 3400i
42
4.6
Saving
The measuring instrument has an internal data memory. It can store up
to 500 data records.
A complete data record consists of:
z Number of the storage location
z Date/time
z Measured value
z Temperature
z Temperature measuring procedure
z ID number
You can transmit measured values (data records) to the data storage
in two ways:
z Save manually
z Switch on AutoStore (Int 1), (see page 44).
Note
If you connect a recorder (analog output), the output on the digital in-
terface is switched off.
4.6.1
Saving manually
You can transmit a measured value to the data storage as follows:
1
Press the <STO> key.
The current number (location number No.) of the next free stor-
age location appears under the current measured value on the
display.
2
Confirm with <RUN/ENTER>.
The display switches to entering the ID number.
Auto
8
%
°F
4
2
°C
TP
mV/pH
mol/L
TempError
RCL
LoBat
AutoCal DIN
CalError
AR
AutoCal TEC ConCal
Time
Baud
Day.Month
Ident
Year
1
U
S
mg/L
% ppm
ISE
2
99
6
pH
4
No.
ARng
Store