0l s – Xylem pH/ION 3400i ION SELECTIVE ELECTRODE ANALYZER User Manual
Page 30

Operation
pH/ 3400i
30
8
Thoroughly rinse the electrode with deionized water.
9
If required, enter the temperature of the second buffer solution
with <
▲> <▼> while keeping the <RUN/ENTER> key de-
pressed.
10
Immerse the pH electrode in the second buffer solution.
11
Press the <RUN/ENTER> key.
12
Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with <
▲> <▼>.
13
When the measured value is stable, press the <RUN/ENTER>
key. The value of the slope (mV/pH) appears on the display.
The probe symbol shows the evaluation of the current calibra-
tion.
14
Press the <RUN/ENTER> key. The value of the asymmetry
(mV) appears on the display.
AutoCal DIN
AutoCal TEC
6