Specifications

9

Input resistance—reference electrode

>1 x 10

10

Ω

Input current—glass electrode

<2 x 10

–12

A

Input current—reference electrode

<1 x 10

–10

A

Measurement range

–1500 to 1500 mV

Measurement error—pH

1

,

2

<0.02 pH units plus sensor error;
TC: 0.002 pH/K

Measurement error—mV

1,2

<1 mV plus sensor error; TC: 0.1 mV/K

Display range—pH

–2.00 to 16.00 pH units

Display range—ORP

–1999 to 1999 mV

pH sensor calibration

Offset range

± 60 mV

Slope range

85 to 103% (47.5 to 61 mV/pH unit)

Calibration timer

0 to 9999 hours

ORP sensor calibration (si792 P, si792x P and si792x P-FF only)

Calibration range

–700 to 700 mV

Calibration timer

0 to 9999 hours

1

(± 1 count plus sensor error)

2

IEC 746 Part 1, at nominal operating conditions