Calibration

71

first buffer solution. When complete, the slope (%) and offset
(mV) of the probe (based on 25 °C) will be displayed.

6. Press

ENTER

,

ENTER

to return to the measurement mode. The

HOLD mode will remain active for 20 seconds.

7. Remove the probes from the buffer, rinse and return to the process.

8.1.1.2

Calibration with manual buffer entry

For calibration with manual buffer specification, the pH value of the
buffer solution at the proper temperature must be entered. This
calibration allows for calibration with any buffer solution.

Be sure that manual calibration is selected in the configure, cal menu
(

section 5.2.3 on page 40

).

1. Press

CAL

, enter passcode: 1100,

ENTER

. The display will show

CAL1.

2. Remove the pH and temperature probes from the process. Clean

the probes and immerse in the first buffer solution (any order).
Briefly stir the solution with the probes.

3. Press

ENTER

. Use the

UP ARROW

and

RIGHT ARROW

to edit the

displayed pH value to match the pH of the buffer solution at the
selected temperature.

4. Press

ENTER

. Measurement of the buffer solution will begin. When

measurement is complete, the display will show CAL2.

5. Complete a one-point calibration or start a two-point calibration:

One-point calibration: press

CAL

to end the calibration. The

slope (%) and offset (mV) of the probe (based on 25 °C) will be
displayed.

Two-point calibration: immerse the probes in the second buffer
solution. Press

ENTER

. Edit the displayed pH value to match the

second buffer solution and press

ENTER

. Measurement of the