Operation—Foundation Fieldbus

43

Select calibration mode / solution

CA.SOL

-01-BUF

Mettler-Toledo

-02-BUF

Merck Titrisols, Riedel Fixanals

-03-BUF

Ciba (94)

-04-BUF

NIST technical buffers

-05-BUF

NIST standard buffers

-06-BUF (default)

HACH buffers

-07-BUF

WTW technical buffers

-08-BUF

Hamilton

MAN (manual entry, for buffers not listed in this table)

CA.timE

CAL timer interval

0000–9999 h (default: 0000 h)

Select Sensocheck

AL.SnSO

CHECK ON

Continuous Sensocheck evaluation of
sensor function

CHECK OFF (default)

LED in Hold mode

AL.LED

HOLD ON

LED blinks during hold

HOLD OFF (default)

LED off during hold

Enter Fieldbus address (optional)

1

FF.ADR

0017–0031 BUS (default: 0026 BUS)

1

Use only when there is no bus connection. The transmitter will restart and set all

parameters to default values. Individual settings must be entered once more.