Operation—foundation fieldbus, 2 configuration menu (continued) – Hach-Lange SI792 P, SI792x P, SI792x P-FF, SI792x P-PA User Manual User Manual
Page 45

Operation—Foundation Fieldbus
43
Select calibration mode / solution
CA.SOL
-01-BUF
Mettler-Toledo
-02-BUF
Merck Titrisols, Riedel Fixanals
-03-BUF
Ciba (94)
-04-BUF
NIST technical buffers
-05-BUF
NIST standard buffers
-06-BUF (default)
HACH buffers
-07-BUF
WTW technical buffers
-08-BUF
Hamilton
MAN (manual entry, for buffers not listed in this table)
CA.timE
CAL timer interval
0000–9999 h (default: 0000 h)
Select Sensocheck
AL.SnSO
CHECK ON
Continuous Sensocheck evaluation of
sensor function
CHECK OFF (default)
LED in Hold mode
AL.LED
HOLD ON
LED blinks during hold
HOLD OFF (default)
LED off during hold
Enter Fieldbus address (optional)
1
FF.ADR
0017–0031 BUS (default: 0026 BUS)
1
Use only when there is no bus connection. The transmitter will restart and set all
parameters to default values. Individual settings must be entered once more.
6.1.2
Configuration menu (continued)