Troubleshooting

83

ERR 13

Current output icon flashes;
reading range value set too
large or too small

Check the range settings
(

Table 7 on page 36

)

ERR 33

Sensocheck icon flashes;
error with glass pH probe

First calibrate the transmitter.
If error repeats check the
following steps.

Make sure of the
configuration settings.

Clean connection terminals
and probe cap.

Replace probe and cable.

ERR 34

Sensocheck icon flashes;
error with reference
electrode

Clean connection terminals
and probe cap.

Replace probe and cable.

Jumper between terminal 4
and 5 is missing
(combination sensors only)

ERR 98

CONF flashes; configuration
or calibration data is
defective. Memory error in
the program.

Return to the factory for
repair and calibration.

ERR 99

FAIL flashes; EEPROM or
RAM defective

Return to the factory for
repair and calibration.

Table 23 Error codes (Continued)

Code

Description

Corrective action