Section 10 Troubleshooting

10.1 Sensoface

The Sensoface feature is active whenever Sensocheck is active. This
feature monitors the electrode for defects in the sensor or cable, and
indicates the maintenance status of the sensor (

Table 21

).

10.2 Sensocheck

The Sensocheck feature monitors the sensor electrodes and wires for
open circuits.

Table 22

shows the probe types that Sensocheck

monitors. When problems are detected, the error message “Err 33” is
displayed, along with an icon corresponding to the type of problem.

Sensocheck can be switched off in the configuration menu.

Note: For confirmation a Sensoface is displayed after calibration.

Note: Set Sensocheck to off when a fast response time is required.

Table 21 Sensoface description

Sensoface

Description

The probe is functioning properly.

The offset and slope of the probe are acceptable, but will
require replacement soon.

The offset and slope of the probe have reached values
which no longer ensure proper calibration or the pH probe
is defective due to a broken glass pH bulb. Replace the
electrode.

Table 22 Sensocheck and electrode types

Electrode type

Measure probe

Reference probe

Sensocheck on

pH/ORP

On

On

pHD

1

1 Not available in EU

Off

Off