3 calibration mode configuration, 4 alarm settings configuration, Section 5.2.3 – Hach-Lange SI792 P, SI792x P, SI792x P-FF, SI792x P-PA User Manual User Manual

Page 42: Section 5.2.4, Operation—4–20 ma/hart

Advertising
background image

Operation—4–20 mA/HART

40

5.2.3

Calibration mode configuration

An overview of the calibration setup menu is shown in

Table 9

. Refer to

Appendix A on page 95

for buffer tables.

5.2.4

Alarm settings configuration

An overview of the alarm setup menu is shown in

Table 10

.

Table 9 Calibration mode setup menu—CAL mnu

Select pH buffer set

CA.SOL

–01–BUF

Knick/Mettler-Toledo

–02–BUF

Merck Titrisols, Riedel Fixanals

–03–BUF

Ciba (94)

–04–BUF

NIST technical buffers

–05–BUF

NIST standard buffers

–06–BUF (default) HACH buffers

–07–BUF

WTW technical buffers

–08–BUF

Hamilton Duracal

MAN (manual entry, for buffers not listed in this table)

DAT—Entry of offset and slope of premeasured electrodes

Enter calibration timer interval

CA.tiME

Enter the time interval for calibration (0 to 9999 hours; default: 0 h
to disable timer)

Table 10 Output setup menu for alarm settings

Select Sensocheck

AL.SnSO

CHECK ON

Continuous Sensocheck evaluation of probe
function

CHECK OFF (default)

Enter alarm delay

AL.dLY

0010 sec (default)

Range: 0000–0600 sec

LED in Hold mode

AL.LED

HOLD ON

LED blinks during hold

HOLD OFF

LED off during hold