1 output configuration, Section 5.2.1, Operation—4–20 ma/hart – Hach-Lange SI792 P, SI792x P, SI792x P-FF, SI792x P-PA User Manual User Manual

Page 38

Advertising
background image

Operation—4–20 mA/HART

36

To exit the configuration mode at any time, press

CONF

and then

ENTER

. The output current will be held for 20 seconds and the

measured value will be displayed.

Note: During configuration the transmitter remains in the Hold mode for
safety reasons. The loop current is held at the value specified in the
o1.HoLD menu option. The Sensoface icon is inactive. The
configuration mode indicator is displayed (

Figure 11 on page 33

).

5.2.1

Output configuration

An overview of the 4–20 mA output setup menu is shown in

Table 7

.

Table 7 Output setup menu—out.1

MNU

Select the parameter

o1.UniT

pH (default) o1.EL (electrode)

Glass (default for EU version)

ISFET

pHD (default for standard ver-
sion; for pHD and LCP sensors)

ORP

Specify the value for the 4 mA signal (

section 5.2.1.1

)

o1.4mA

Enter the pH value (–2.00 to 16.00 pH; default: 0 pH)

Enter the ORP value (–1999 to 1999 mV; default: 0 mV)

Specify the value for the 20 mA signal (

section 5.2.1.1

)

o1.20mA

Enter the pH value (–2.00 to 16.00 pH; default: 14 pH)

Enter the ORP value (–1999 to 1999 mV; default: 14 mV)

Set time averaging filter for reducing noise (

section 5.2.1.2

)

o1.Ftme

Enter the time in seconds (0 to 120 seconds; default: 0 seconds)

Send a 22 mA signal during errors (

section 5.2.1.3

)

o1.FAil

On

Off (default)

Specify the value to output during HOLD periods (

section 5.2.1.4

)

o1.HoLD

LAST measured value (default)

FIXed

Enter the fixed value to output in mA (3.8 to 22 mA;
default: 21 mA)