2 temperature compensation configuration, Section 5.2.2, Operation—4–20 ma/hart – Hach-Lange SI792 P, SI792x P, SI792x P-FF, SI792x P-PA User Manual User Manual

Page 41

Advertising
background image

Operation—4–20 mA/HART

39

5.2.2

Temperature compensation configuration

An overview of the temperature compensation setup menu is shown in

Table 8

.

Press

ENTER

to access a menu item. Use the

ARROW KEYS

to edit

values. Press

ENTER

to save the settings. If a value is outside of the

acceptable range, “Err” will be displayed and the value will not be
accepted. To exit the menu and return to the measurement mode, press

CONF

and then

ENTER

.

Table 8 Temperature compensation setup menu—tc

MNU

Select temperature unit

tc.UnIT

°C (default)

°F

Select temperature probe

tc.rTD

Pt100

Pt1000 (default for EU version)

NTC300 (default for standard version with pHD and LCP sensors)

Temperature detection during measurement

tc.MEAS

Auto (default)

Manual

Enter the temperature (–20 to 200 °C; default: 25 °C)

Enter the temperature (–4.0 to 392 °F; default: 25 °F)

Temperature detection during calibration

tc.CAL

Auto (default)

Manual

Enter the temperature (–20 to 200 °C; default: 25 °C)

Enter the temperature (–4 to 392 °F; default: 25 °F)

pH only temperature compensation of process medium

tc.LIN

Enter compensation (–19.99 to 19.99%/K; default: 0%/K)