Calibration

72

buffer solution will begin. When complete, the slope (%) and
offset (mV) of the probe (based on 25 °C) will be displayed.

6. Press

ENTER

,

ENTER

to return to the measurement mode. The

HOLD mode will remain active for 20 seconds.

7. Remove the probes from the buffer, rinse and return to the process.

8.1.1.3

Enter calibration data from calibrated probes

If the slope and offset of a probe have been determined from a prior
calibration, these values can be entered into the transmitter.

Be sure that DAT is selected in the configure, cal menu (

section 5.2.3

on page 40

).

1. Press

CAL

, enter passcode: 1100,

ENTER

. The display will show

CAL and then a screen for offset (mV) entry.

2. Enter the offset mV value of the probe. Press

ENTER

. The value of

the slope (%) will be displayed.

3. Enter the slope value of the probe in percent (%).

Table 19

shows

the slope in percent vs. mV/pH unit for reference. Press

ENTER

. The

slope (%) and offset (mV) of the probe will be displayed.

4. Press

ENTER

,

ENTER

to return to the measurement mode. The

HOLD mode will remain active for 20 seconds.

Table 19 Slope in % vs. mV/pH unit

Slope (%)

Slope (mV/pH)

78

46.2

80

47.4

82

48.5

84

49.7

86

50.9

88

52.1

90

53.3

92

54.5